In the event that something goes wrong with the instrument, carry out the measures below.
If the problem persists, a breakdown of the instrument may have occurred. Contact the
nearest KONICA MINOLTA-authorized service facility. When contacting us, please tell us
the malfunction No. and the version of your instrument. For the procedure to check the
version of the instrument, refer to page 62.
Malfunction
No.
Condition Cause Action
Reference
page
1
The instrument is
turned on but the
LCD is blank.
Is the AC adapter connected to
an AC outlet?
Connect the AC adapter.
22
Is the AC adapter connected to
the instrument?
Connect the AC adapter.
22
Is a proper AC adapter
connected?
Only use the optional AC
adapter specied by KONICA
MINOLTA (AC-A305J/L/M).
9
Is the AC voltage within the
range of the rating?
Use the instrument with a
voltage within the range of
nominal voltage ± 10%.
9
Are the batteries loaded? Load the batteries.
21
Is the battery voltage low? Replace the batteries with new
ones. Or, connect the AC
adapter.
21
2
I looked into the
viewnder but the
eld is too dark to
see anything.
Is the lens cap still on the
objective lens?
Remove the lens cap.
-
Is the eyepiece ND lter
attached to the viewnder?
Use the eyepiece ND lter only
when measuring a bright
object.
-
3
Key operation is
not accepted.
Is the instrument set in remote
mode?
Press the
ESC
key to cancel
remote mode.
95
Is there any possibility that a
key that does not work was
pressed?
Press a correct key.
-
4
When I enter a
calibration value/
target, the value
displayed after
the setting is
different from the
entered value.
The value might be different by
1 digit due to calculation error.
108
Checking for Malfunction
105
Explanation and
Information
ERROR
CHECK