Chapter 2 Before the Repair
2-5
2.3 Principal Specifications
2.3.1 Equipment Specification
Applicable cassette
Dimensions 740 (W) x 365 (D) x 747 (H) mm
Weight Approximately 100 kg
Floor height adjustment range Up to 12 mm bump at the position of the foot adjuster
Noise level During operation: 55 dB or less
During idle: 42 dB or less
(Excluding warning buzzer, user operating sound, etc.)
Power consumption 185 Wh or less (40 mR or less exposure, continuous processing of
14 x 17 inches)
Maximum power consumption 800 VA or less (100 V, 8 A or less)
Power supply In Japan AC 100 V ± 10% (50/60 Hz)
North America AC 110/115/120/220/230/240 V ± 10%
(60 Hz)
Europe AC 200/220/230/240 V ± 10% (50 Hz)
Environment condition
(Temperature/Humidity)
During operation: 15 ~ 30°C/35 ~ 80% RH (no condensation)
Not operating: -10 ~ 40°C/10 ~ 95% RH (no condensation)
Periodic replacement parts Erase lamp unit (erase count: About 30,000 shots)
Accessories Power supply cable, operation manual
Available accessories REGIUS cassette, REGIUS plate, JM, REGIUS console (CS-1/CS-
2/CS-3)
Cassette type Description
Readable cassette type RC-110, RC-110R, and cassette with equivalent open/close/
separation mechanism with RC-110
Readable cassette size 14 x 17 inches, 14 x 14 inches, 11 x 14 inches,
10 x 12 inches, 8 x 10 inches, 24 x 30 cm, 18 x 24 cm,
15 x 30 cm
Exposure only cassette type RC-110T (long length exposure), RC-110L (confirming therapy
positioning)
Exposure only cassette size RC-110T: 4 types (10 x 36 inches, 11 x 28 inches, 14 x 42 inches,
14 x 51 inches)
RC-110L: 3 types (14 x 17 inches, 14 x 14 inches, 10 x 12 inches)
Im
ortant
Cassettes produced before April 2004 can not be used.