Chapter 3 Troubleshooting
3-13
References for response method
•
See "2.5.3 Transporter Unit (Page 2-13)" for the positions of push plate and tumbler.
• See "2.5.8 Framework (Page 2-19)" for the position of the transporter motor, the transporter read
sensor, the transporter HP sensor, and the MDU.
• See "7.2.8 Greasing Transporter Lead Screw Unit and Transporter Motor Unit Assembly Gear
Mechanism Unit (Page 7-12)" for applying grease to the transporter motor unit.
• See "5.10.5 Replacing the Transporter Read Sensor (Page 5-122)" for the procedure to replace the
transporter read sensor.
• See "5.10.4 Replacing the Transporter HP Sensor (Page 5-121)" for the procedure to replace the
transporter HP sensor.
• See "5.10.6 Replacing the Transporter Motor (Page 5-123)" for the procedure to replace the
transporter motor.
• See "5.10.2 Replacing the MDU (Page 5-116)" for the procedure to replace the MDU.
3.2.4 06150 ~ 06182: Subscan Motor Errors
• Error codes S column: Errors at the read operation/calibration.
• Error codes E column: Errors during erase operation.
• Error codes I column: Errors during initialization.
06113 06128 06143 Transporter motor control
error (when moving to HP 2)
There was a trouble with the
motor control when the push
plate unit was moving from the
HP pass position to the home
position, when it was trying to
move from the press down
position to the home position.
1. Confirm the ground. (It may cause
noise)
2. Confirm that the antistatic brush right
under the insertion slot has not fallen off.
3. Confirm following and replace the
faulty parts.
• Ground wire of transporter unit
(if it is not broken)
• Transporter motor
• Motor wire:
JP52 ~ MCN6 (MDU)
• MDU
Error code Description of the error Response
S E I
06150 06165 06180 Encoder initialization error
Initialization of the encoder has
failed.
1. Confirm that the fixing shaft of the
simple fixing lock is not lowered.
2. Confirm there are no foreign objects
on the magnet shaft of the subscan
unit, and remove it if there are any.
3. Confirm that it is installed horizontally.
4. Confirm following and replace the
faulty parts.
• Encoder
• Encoder wire: CN4 (LMC)
• LMC
Error code Description of the error Response
(References to replacement methods are
listed at the end of the table.)
S E I

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta REGIUS 110 and is the answer not in the manual?

Konica Minolta REGIUS 110 Specifications

General IconGeneral
BrandKonica Minolta
ModelREGIUS 110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals