Chapter 7 Periodic Maintenance
7-13
7.2.9 Greasing of Release Shaft Holder
Apply grease to the release shaft holder.
1 Apply grease to the release shaft holder.
•
Grease: Plusguard No. 2 by Kyodo Yushi
This completes the greasing procedure for the release shaft holder.
Go to "7.2.10 Greasing Justifier Motor Unit Assembly Gear Mechanism (Page 7-13)" and apply grease to
the justifier motor unit assembly gear mechanism.
7.2.10 Greasing Justifier Motor Unit Assembly Gear Mechanism
Apply grease to the justifier motor unit assembly gear mechanism unit.
1 Remove the exterior frame (front).
•
4 screws (M4 x 8)
Loosen all screws.
2 Apply grease to the justifier motor unit assembly
gear mechanism unit.
•
Grease: Plusguard No. 2 by Kyodo Yushi
Apply grease on entire area by turning A.
3 Attach the exterior frame (front) that was
removed in Step 1.
•
4 screws (M4 x 8)
Caution
Be sure to turn OFF the power supply circuit breaker and unplug the power cable from the
equipment whenever the maintenance is performed.
Application Point
Im
p
ortant
Always use the specified grease.
Caution
Be sure to turn OFF the power supply circuit breaker and unplug the power cable from the
equipment whenever the maintenance is performed.
First Front Side
A
First Front Side
Im
p
ortant
Always use the specified grease.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta REGIUS 110 and is the answer not in the manual?

Konica Minolta REGIUS 110 Specifications

General IconGeneral
BrandKonica Minolta
ModelREGIUS 110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals