Chapter 5 Disassembly and Assembly
5-55
5.4.10 Replacing the Receiver (Receiver Unit Assembly)
This section describes the instructions on how to replace the receiver (receiver unit assembly).
Work outline
Listed below are the number of people and the rough standard of work hours required for performing the
works described in this section.
Requirements
Shown below are the tools requiring to be prepared in addition to the standard tools before performing the
works in this section.
1 See Step 1 to Step 6 in "5.4.12 Replacing the Receiver Belt (Page 5-59)" to remove the
receiver belt.
2 Remove the connector (JP17) from the relay
cable on the receiver sensor.
3 Cut off the wiring band (at 2 locations) to lead the
cable inside the equipment.
4 Remove the receiver.
•
4 screws (M4 x 6)
Caution
Be sure to turn off the power supply circuit breaker and unplug the power cable from the
equipment whenever you disassemble/assemble the equipment.
Personnel
Number
Work Hours (steps 2 to 7)
1
No. Tool
1 Wiring band
JP17
First Front Side
Caution
When removing the connector, hold a notch
on the connector without pulling a cable.
Giving a strong pull to the cable may cause
it to be broken.
Caution
The receiver is secured to the guide. Work
must be performed, supporting or setting
down the guide so as to keep it up.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta REGIUS 110 and is the answer not in the manual?

Konica Minolta REGIUS 110 Specifications

General IconGeneral
BrandKonica Minolta
ModelREGIUS 110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals