Chapter 7 Periodic Maintenance
7-10
7.2.6 Cleaning Cassette Absorption Magnet
Clean the surface of the magnet on the cassette absorption plate.Confirm that the surface of the cassette
absorption plate is free of scratches and flat at the same time.
1 Open the second front door.
2 Wipe the surface (magnet face) of the cassette
absorption plate lightly with alcohol.
Confirm that it is free of scratches and flat at the
same time.
This completes the cleaning of the cassette absorption magnet.
Go to "7.2.7 Greasing LM Guide (Page 7-11)" and grease the LM guide.
Caution
Be sure to turn OFF the power supply circuit breaker and unplug the power cable from the
equipment whenever the maintenance is performed.
Caution
Be careful not to distort the cassette
absorption plate with too much force.
Im
p
ortant
Replacement of the transporter unit (at the
factory) will be necessary if there is a
scratch on the surface of the cassette
absorption plate or if it is not flat.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta REGIUS 110 and is the answer not in the manual?

Konica Minolta REGIUS 110 Specifications

General IconGeneral
BrandKonica Minolta
ModelREGIUS 110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals