EasyManuals Logo
Home>Konica Minolta>Measuring Instruments>REGIUS 110

Konica Minolta REGIUS 110 Service Manual

Konica Minolta REGIUS 110
302 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
Chapter 3 Troubleshooting
3-42
Phenomenon: 1 to few lines of horizontal line looking digital
3.4.5 Horizontal Streak All Over the Image
Phenomenon: Horizontal streak in 7 line cycle
Phenomenon: Irregular horizontal streaks all over the left side of the image
Phenomenon: White streaks starting in the midway going all the way to the right edge of
the image
Phenomenon: Horizontal streak in 6 ~ 7 mm cycle
No. Cause Response
1 Connection failure Confirm the connection of the LMC connectors (CN8,
J1, CN301, CN302, CN303, and CN304).
2 H-sync signal detection problem. Confirm the H-sync signal using the oscilloscope and
replace the optical unit if there is a fault.
• See "3.2.19 Confirm the H-sync Signal (Page 3-
30)" for the procedure to confirm the H-sync.
• See "5.2.5 Removing/Installing the Optical Unit
(Page 5-11)" for the procedure to replace the
optical unit.
3 When there is any other digitalization
(include density change)
Confirm after replacing LMC or CIU.
• Replace the LMC.
See "5.6.2 Replacing the LMC (Page 5-79)" for
the procedure to replace the LMC.
• Replace the CIU.
See "5.10.1 Replacing the CIU (Page 5-115)" for
the procedure to replace the CIU.
Im
p
ortant
Always perform the image size adjustment, the image position adjustment, the unevenness
calibration, and the sensitivity calibration after replacing the optical unit.
No. Cause Response
1 Polygon correction Redo the unevenness calibration.
2 There is a possibility that vignetting is
occurring to the laser ray in the optical
unit
Replace the optical unit.
See "5.2.5 Removing/Installing the Optical Unit (Page
5-11)" for the procedure to replace the optical unit.
No. Cause Response
1 Fluctuation of laser intensity Replace the optical unit.
See "5.2.5 Removing/Installing the Optical Unit (Page
5-11)" for the procedure to replace the optical unit.
No. Cause Response
1 Fluctuation of laser intensity Replace the optical unit.
See "5.2.5 Removing/Installing the Optical Unit (Page
5-11)" for the procedure to replace the optical unit.
No. Cause Response
1 Problem in subscan action due to the
rust on the LM guide
Lubricate the LM guide.
See "7.2.7 Greasing LM Guide (Page 7-11)" for the
procedure to lubricate the LM guide.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta REGIUS 110 and is the answer not in the manual?

Konica Minolta REGIUS 110 Specifications

General IconGeneral
BrandKonica Minolta
ModelREGIUS 110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals