Chapter 6 Adjustment
6-4
6.2 Adjust Receiver Belt Tension
Adjust the tension of the timing belt for the receiver mechanism in the receiver unit.
Work outline
Listed below are the number of people and the rough standard of work hours required for performing the
works described in this section.
Requirements
Shown below are the tools requiring to be prepared in addition to the standard tools before performing the
works in this section.
Positioning and adjustment parts for timing belt
Adjustment method
1 See "îš„ Removal Procedures (Page 5-6)" in "5.2.3 Removing/Installing the Exterior Panel and
Insertion Unit" to remove the first front panel.
2 Remove the electric component mount panel.
•
8 hex/Phillips-head screws (M4 x 8)
Loosen all screws.
Personnel
Number
Work Hours (steps 2 to 7)
1
No. Tool
1 Acoustic wave tension meter
Caution
Be sure to turn off the power supply circuit breaker and unplug the power cable from the
equipment whenever you adjust the unit position of the equipment.
Timing Belt
Receiver Idler Pulley Unit Assembly
First Front Side

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta REGIUS 110 and is the answer not in the manual?

Konica Minolta REGIUS 110 Specifications

General IconGeneral
BrandKonica Minolta
ModelREGIUS 110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals