Operator’s Manual of Anesthesia System 6 - 5
Operations Set Ventilation
NOTE: You do not need to perform this operation if inspiratory anesthetic
agent is not used.
NOTE: The anesthesia system can be installed with vaporizers of Isoflurane,
Sevoflurane and Desflurane. Only one vaporizer can be opened at a time.
6.5.2.1 Select the Desired Anesthetic Agent
1. Determine the anesthetic agent to be used and then fill the vaporizer.
NOTE: Only vaporizers compliant with ISO 80601-2-13 and with Selectatec
Interlock-Systems may be Installed to this equipment. Refer to the
vaporizer manufacturer’s instructions for filling, draining and other
information of the vaporizer.
WARNING: Ensure that the correct anesthetic agent is used. The vaporizer is
designed with the specific anesthetic agent named on it and further
indicated by color coded labeling. The actual output concentration of the
anesthetic agent will vary if the vaporizer is filled with the wrong agent.
2. Install the vaporizer filled with anesthetic agent onto the anesthesia system.
6.5.2.2 Adjust the Concentration of Anesthetic Agent
Push and turn the concentration control on the vaporizer to set the appropriate concentration of
anesthetic agent. For details about how to use the anesthetic agent, refer to the Vaporizer
Instructions for Use.
6.6 Set Ventilation
NOTE: In all ventilation modes, when inspiration pressure reaches the upper
limit of PAW alarms, the system switches to expiration immediately and
airway pressure is released.
NOTE: In case of a fault with the drive gas supply, the Auto ventilation mode
won't function normally.
6.6.1 Change Ventilation Mode
To change ventilation mode to Manual:
Use the Auto/Manual switch on the breathing system to enter and exit Manual ventilation mode.
To change ventilation mode to VCV, SIMV-VC, PCV, PCV-VG, SIMV-PC, SIMV-VG, CPAP/PS,
APRV:
1. Select the tab of the desired ventilation mode. The [Set Mode] button (or [Preset Mode] button
in manual mode) will flash (Figure 6-2).
2. Select the [Set Mode] button (or [Preset Mode] button in manual mode) to confirm the
change. If the [Set Mode] button is not selected after several seconds, an audio reminder will
sound for several seconds and then the system will return to the previous ventilation mode.
3. Optionally, select each available ventilation parameter to edit the parameter setting.
4. Move the Auto/Manual Switch to the Auto position.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
BrandMindray
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals