Ventilation Linkage Parameters Parameters and Factory Defaults
B - 10 Operator’s Manual of Anesthesia System
B.11 Ventilation Linkage Parameters
The table below describes how ventilation mode changes impact parameter values. For example, when the ventilation mode changes, its parameter values may
be shared by the other ventilation mode with the same parameters. When the ventilation mode changes, other parameters may have different values set.
CURRENT VENTILATION
MODE & AFFECTED
PARAMETERS
PREVIOUS VENTILATION MODE
VCV SIMV-VC PCV PCV-VG SIMV-PC CPAP/PS SIMV-VG APRV
VCV Vt —*Vti mea-
surement or
the closest
approxima-
tion
*##*#
RR —** * *# *#
I:E —#* * ## ##
Tpause —*# # ## ##
PEEP —** * ** *#
Plimit —*# * ## *#
SIMV-VC Vt *—Vti mea-
surement or
the closest
approxima-
tion
*##*#
RR *—* **#*#
Tinsp #—# #*#*#
Tpause *—# #####
PEEP *—* ****#
Plimit *—# *##*#
ΔPsupp #—# #***#
F-Trig/P-Trig #—# #***#
Exp% #—# #***#
Trig Window #—# #*#*#
Tslope #—* *****
* In the event that the previous ventilation mode and the current one share some parameter values, the parameter setting remains unchanged.
# For the parameters that are not shared by the previous ventilation mode and the current one, the most recent settings of the parameters apply.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
BrandMindray
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals