System Soft Key System Interface
4 - 26 Operator’s Manual of Anesthesia System
4.6.1.5 Export
Insert a U disk to the USB interface of the equipment as per the prompts on the screen. Select the
[Export] soft key to export list trends, graphic trends, event logs and captured screens to the U disk.
The exported data is in the format of .html. Files in the format can be opened in Internet Explorer 8.0,
9.0, 10.0 and 11.0.
4.6.2 Lung Recruitment (Optional)
See (Pages 6-14) 6.6.10 "Lung Recruitment Ventilation (Optional)".
4.6.3 Flow Pause
Use [Flow Pause] to temporarily suspend fresh gas flow during the ventilation. Using [Flow Pause]
while the breathing system is disconnected from patient prevents the flow of gas into the room. [Flow
Pause] is available during both mechanical ventilation and manual ventilation.
To enter the [Flow Pause] state:
1. On the Main Screen, select the [Flow Pause] soft key.
2. Select [Ye s ] on the pop-up screen to confirm the change. The system will enter the [Flow
Pause] state.
T
Figure 4-30 Flow pause
When the system is in the [Flow Pause] state:
• The fresh gas flow is turned off.
• The mechanical ventilation is suspended.
• Physiological alarms related to ventilation and gas are disabled.
• The countdown timer is enabled. The default countdown time is 60 seconds. You can select [+30 sec] button
to add 30 seconds to the current countdown time. The maximum countdown time is 2 minutes.
To exit the [Flow Pause] state:
• The system exits the [Flow Pause] state automatically when the countdown time is 00:00.
• Select [End Flow Pause] button to exit the [Flow Pause] state.
• The system exits the [Flow Pause] state automatically when the system enters Standby mode or the BFCS is
enabled.
After the system exits from the [Flow Pause] state: