Operator’s Manual of Anesthesia System 4 - 5
System Interface System Information Title
4.2.1.1 Set Patient Information in Anesthesia System
NOTE: The patient type can be changed only when the equipment is in the
standby or manual mode.
1. Select the Patient Information icon to open the Patient Information settings menu.
2. Set the patient information.
3. Select to confirm the change and close the menu.
4.2.1.2 Retrieve Patient Information from ADT Server
The anesthesia system can be connected to the ADT server through eGateway to download patient
information from the ADT server.
1. Connect the network cable.
2. Set the network and ensure that the network access is normal.
a. Enter Standby mode.
b. Select the icon to open the [Setup] menu.
c. Select the [System] soft key, enter the system password and confirm the password.
d. Select the [Network] tab in the system menu.
e. Select [ADT] and set the [ADT] to (on) on the pop-up screen.
f. Set [Destination IP] and [Port].
g. Select the [Te st ] soft key.
h. Confirm that the test result is [Connected].
3. Select the [Find Patient] soft key in the Patient Information menu.
4. Enter the key information on the pop-up screen.
5. Select the [Search] soft key and a list of conforming patient information will be displayed on the
screen.
6. Select the desired patient information in the list and select the [Import] soft key. The imported
data includes patient ID, visit number, first name, last name, bed, room, department and facility.
4.2.1.3 Synchronize Patient Information
With the anesthesia system connected to the eGateway, when key words are entered in the
anesthesia system and eGateway recognizes the patient information related to the key words, it will
update the patient information to the anesthesia system synchronously. If the patient information in
the eGateway is updated, eGateway will also update the information to the anesthesia system
synchronously. Synchronizable patient information include: patient ID, visit number, last name,
first name, date of birth, age, weight, height and gender.
1. Connect the network cable.
Bed
Enter up to 30 digits for each field.
Room
Department
Facility
EDITABLE FIELD DESCRIPTION
Table 4-2 Patient information

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
BrandMindray
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals