Operator’s Manual of Anesthesia System 7 - 9
Anesthetic Gases and O
2
Concentration Monitoring Measurement Limits
3. Enter the desired parameter value using the keyboard on the screen. For each parameter,
the range of values is displayed above the keypad.
4. You can select the [Load Alarm Defaults] button to restore the default values. This restores the
high and low limits of parameters to the user default values.
5. Select to save the changes (or select to discard the changes).
Figure 7-5 [Agents] tab
7.8 Measurement Limits
The following factors may reduce measurement accuracy:
Leakage or internal sample gas leakage
Mechanical shock
Humidity or condensate
Cyclic pressure greater than 10 kPa (100 cmH
2
O)
Other interference source (if present)
NOTE: The output gas data of the anesthesia module will be zero if the
measured concentration is below the defined threshold for more than
3seconds.