Other Features Operations
6 - 18 Operator’s Manual of Anesthesia System
Preset Ventilation Mode after Exit from ACGO
For example, if the current ventilation mode is VCV, enable ACGO and the system will enter the ACGO
mode. Under such circumstances:
• If the ventilation mode remains unchanged, the system will enter the VCV mode upon its exit
from the ACGO mode.
• If another ventilation mode other than [VCV] has been selected, such as [PCV], select [Preset
Mode], and the system will enter the PCV mode upon exit from the ACGO mode.
NOTE: If the system is currently in Standby or Manual mode, the system enters
ACGO mode when ACGO is enabled. Under such circumstances, the
[Preset Mode] feature is invalid. At this point, the system will enter the
corresponding Standby or Manual mode upon exit from the ACGO mode.
Figure 6-15 Press the [Preset Mode] key to set [PCV] as the system's ventilation mode upon exit
from the ACGO mode
6.7.2 Monitor
When the anesthesia system has an external AG module configured, the system supports the
Monitor mode. In the Manual ventilation mode, select the [Monitor Mode] soft key to enable the
Monitor mode. [Monitor] is displayed in the Current Mode area. The mode disables all ventilation-
related alarms.
Figure 6-16 Monitor Mode

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
BrandMindray
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals