Ventilator Specifications Product Specifications
10 - 22 Operator’s Manual of Anesthesia System
Thigh: 0.2 to 10 s, Step: 0.1 s
Tlow: 0.2 to 10 s, Step: 0.1 s
Exp% 5 to 80 %, Step: 1 %
Ventilator Monitored Parameters Range
PEAK: -20 to 120 cmH
2
O
PLAT: -20 to 120 cmH
2
O
MEAN: -20 to 120 cmH
2
O
PEEP: 0 to 70 cmH
2
O
Tidal Volume (Vt/Vt insp) (Under BTPS
condition):
0 to 3000 ml
Minute Volume (MV/MV insp) (Under
BTPS condition):
0 to 100 L/min
RR: 0 to 150 bpm
I:E: 4:1 to 1:10
Raw: 0 to 600 cmH
2
O/(L/s)
Compl: 0 to 300 mL/cmH
2
O
MVLeak (Under BTPS condition): 0 to 10 L/min
TABLE 10-32
Ventilator Monitored Parameters and Range
Control Parameters Accuracy
Vt (VCV, SIMV-VC, under BTPS
condition):
10 to 60 ml: ± 10 ml
60 to 210 ml (not including 60 ml): ± 15 ml
210 to 2000 ml (not including 210 ml): ± 7 % of the setting
value
Vt (PCV-VG, SIMV-VG, under BTPS
condition):
5 to 60 ml: ± 10 ml
60 to 210 ml (not including 60 ml): ± 15 ml
210 to 2000 ml (not including 210 ml): ± 7 % of the setting
value
Pinsp: ± 2.0 cmH
2
O or ± 7 % of the setting value, whichever is
greater, but is not more than ±5 cmH
2
O
ΔPinsp:
ΔPsupp: ± 2.0 cmH
2
O or ± 7 % of the setting value, whichever is
greater
ΔPapnea:
Plimit: ±10% of the setting value
PEEP ± 2.0 cmH
2
O or ± 7 % of the setting value, whichever is
greater
Tslope: ± 0.2 s or ± 20 % of the setting value, whichever is greater
RR: ± 1 bpm or ± 10 % of the setting value, whichever is smaller
Min RR:
I:E: 2:1 to 1:4: ±10% of the setting value
Other range: ±25% of the setting value
Apnea I:E:
Apnea Tinsp: ± 0.2 s
Tinsp:
TABLE 10-33 Ventilator Control Accuracy
TABLE 10-31 Ventilator Setting Parameter and Range