Operator’s Manual of Anesthesia System 1 - 9
Safety Safety Information
CAUTION: Do not perform soaking or high-temperature processing on the
O
2
sensor.
CAUTION: Users should monitor oxygen percentage (FiO
2
%) when using the Auxiliary
O
2
/Air Flow Meters. Without oxygen monitoring, it would be impossible to
know the concentration of oxygen delivered to the patient.
CAUTION: The equipment is NOT suitable for use in a magnetic resonance imaging
(MRI) environment.
CAUTION: To ensure measurement accuracy and to avoid possible damage to the
equipment, use only Mindray-approved cables and accessories.
CAUTION: Use the power cord provided with the product. If a substitute is
necessary, use power cord in compliance with the specification.
CAUTION: Do not use a damaged device or accessory. Periodically check all cables
(e.g., AC line cord and patient connection cables) for damage that may
occur through normal use. Replace cables if damaged in any way.
CAUTION: Use of other oxygen sensors may cause improper oximeter performance.
CAUTION: Unintended movement may occur if the casters are not locked.
The operator should lock casters during use of the equipment.
CAUTION: Unsecured devices may slide off the top shelf. Devices should be
securely attached to the top shelf.
CAUTION: The voltage on the auxiliary outlets should be the same voltage as the
outlet into which the equipment is plugged. Ensure that devices
plugged into the auxiliary outlets are rated for the same supply voltage
as the equipment.
CAUTION: During the transport and storage of the vaporizer, block the gas inlet
and outlet of the vaporizer with plugs to prevent foreign substances
from entering the vaporizer.
CAUTION: Do not use any flow outlets as handles when moving the equipment.
The flow outlets may become damaged. Use the metal side bars on the
main body when moving the equipment.
CAUTION: Do not push down on the bag arm forcefully or hang heavy objects
onto it. Excessive weight may bend and damage the bag arm.
CAUTION: Use caution when disconnecting [quick connectors], as the sudden
release of pressure may cause injury.
CAUTION: Avoid factors that can contribute to deterioration of the hose
assemblies. Factors include excessive bending, crushing, abrasion,
system pressures and temperatures that exceed hose ratings, and
improper installation.
CAUTION: Be careful in lifting and manipulating the breathing system during
disassembly of the system.
CAUTION: When the electronic flow control system is disabled, the backup flow
control system will be enabled. The initial flow of backup flow control
system is 1 L/min of O
2
. The backup flow control system display only
has a total flowmeter which can display a maximum flow of 15 L/min.