Installation and Disassembly
3 - 2 Operator’s Manual of Anesthesia System
WARNING: Before installation, check the adequacy of the surface of the structure to
which the equipment is to be attached.
WARNING: This equipment must be installed by a factory authorized
representative.
WARNING: Continuous use of desiccated soda lime may endanger patient safety.
Adequate precautions should be taken to ensure that the soda lime in
the CO
2
absorbent canister does not become desiccated. Turn off all
gases when finished using the system.
WARNING: When electrosurgical equipment is used, keep the electrosurgical leads
away from the breathing system, the O
2
sensor, and other parts of the
anesthesia system. Keep available backup manual ventilation and a
respirator with mask in case the electrosurgical equipment prevents
safe use of the ventilator. Ensure the correct operations of all life
support and monitoring equipment.
WARNING: Do not use masks or breathing tubes that are antistatic or conductive.
They can cause burns if they are used near high frequency
electrosurgical equipment.
WARNING: This anesthesia system has waste gas exhaust ports. The operator of
the machine should pay attention to the disposal of the residual
breathing gas scavenged.
CAUTION: The operational environment and the power source of the equipment
must comply with the requirements as specified in the "Product
Specifications" on Pages 10-1.