3-2
3.2.1 Unpacking
1. When the A8/A9 is received, immediately inspect the box for any damage.
a. If there is no damage and all tip indicators on the box exterior are intact, sign and date the bill
of lading or airway bill to indicate safe receipt of the anesthesia machine.
b. If there is damage or any of the tip indicators on the box exterior have activated, conditionally
accept the delivery and clearly describe the damages on the bill of lading or airway bill. Both the
carrier and recipient must sign and date the bill of lading or airway bill. Save all damaged factory
packaging until further instructed by Mindray. The receiver should immediately contact Mindray
Customer Service.
2. Cut, remove, and discard the white shipping straps from the box.
3. Pull the box top straight up off the box and place it on the floor near the machine. The box
top will be used later as a ramp when rolling down the A8/A9 onto the floor.
4. Remove the protection carton from the top of the machine.
Top plate