8-35
8.38.1.2 Preparations
Before disassembly,
Make sure that the anesthesia machine is turned off and disconnected from the A/C power
source.
Maneuver the anesthesia machine to an appropriate location and then apply the brake.
8.38.1.3 Pre-disassembly
1. Refer to 8.1.4Remove the Work Surface Cover Plate FRU to remove the work surface
cover plate.
2. Refer to step 1 in 8.23.2Remove the Patient Circuit Assembly to remove the patient circuit.
3. Refer to 8.28Disassemble the Volume Exchanger Assembly to remove the volume
exchanger assembly.
4. Refer to 8.29Disassemble the VE Box Assembly to remove the VE box assembly.
8.38.2 Remove the Drive Gas Assembly
1. Remove the seven screws with the Phillips screwdriver.
2. Disconnect the hose and cable from the drive gas assembly and remove the assembly.
8.39 Disassemble the ACGO Assembly
8.39.1 Prepare for Disassembly
8.39.1.1 Tools
During parts disassembly and replacement, the following tools may be required:
Phillips screwdriver
8.39.1.2 Preparations
Before disassembly,
Make sure that the anesthesia machine is turned off and disconnected from the A/C power
source.
Maneuver the anesthesia machine to an appropriate location and then apply the brake.
8.39.1.3 Pre-disassembly
Remove the drive gas assembly (without disconnecting the hose or cable). Refer to step 1 in
section 8.38.2.
8.39.2 Remove the ACGO Assembly
1. Disconnect the hose from the ACGO assembly and remove the three screws with the
Phillips screwdriver.
2. Move the drive gas assembly forward along the arrow direction, to reserve sufficient space
for removing the ACGO assembly.