11-5
11.1.4 High-Flow O2 Supply Assembly, Anesthesia Calculation Inlet
Pipeline Assembly, and EFCS Flowmeter Service Part
No. Name P/N Test Item After
Replacement
Remarks
1
High-Flow O2 Supply
Assembly
115-077955-00
5.9.1 High-flow O2
supply test
Power on the
A8/A9 and
switch the
HFNC on
after the
maintenance.
2
Anesthesia calculation
inlet pipeline assembly
(A9)
115-065016-00
5.9.5 Internal AG module
test; 5.7.1Vaporizer
Monitoring Function Test
/
3
EFCS flowmeter FRU
(O2/N2O/Air)
115-077959-00
5.5 EFCS accuracy test
5.6 BFCS function test