8.41 Disassemble the Air Filter
8.41.1 Prepare for Disassembly
8.41.1.1 Tools
During parts disassembly and replacement, the following tools may be required:
Phillips screwdriver
8.41.1.2 Preparations
Before disassembly,
Make sure that the anesthesia machine is turned off and disconnected from the A/C power
source.
Maneuver the anesthesia machine to an appropriate location and then apply the brake.
8.41.1.3 Pre-disassembly
1. Refer to 8.1.4Remove the Work Surface Cover Plate FRU to remove the work surface
cover plate.
2. Refer to step 1 in 8.23.2Remove the Patient Circuit Assembly to remove the patient circuit.
3. Refer to 8.28Disassemble the Volume Exchanger Assembly to remove the volume
exchanger assembly.
4. Refer to 8.29Disassemble the VE Box Assembly to remove the VE box assembly.
8.41.2 Remove the Air Filter
Pull out the air filter.
8.42 Disassemble the AG Module
8.42.1 Prepare for Disassembly
8.42.1.1 Tools
During parts disassembly and replacement, the following tools may be required:
Phillips screwdriver
8.42.1.2 Preparations
Before disassembly,
Make sure that the anesthesia machine is turned off and disconnected from the A/C power
source.
Maneuver the anesthesia machine to an appropriate location and then apply the brake.