EasyManuals Logo
Home>Mindray>Medical Equipment>A9

Mindray A9 Service Manual

Mindray A9
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #396 background imageLoading...
Page #396 background image
11-12
11.1.9 Drawer and Battery Box Assemblies
No. Name P/N Test Item After
Replacement
Remarks
1
Battery box
assembly (two
batteries)
115-066846-00
Power on the machine, access
the diagnostic maintenance
screen, select Data
Monitors > Power System,
and check the battery quantity
and voltage.
Includes two
batteries.
2
Upper-middle
drawer assembly
service part
115-066841-00 /
The drawer
assembly and
the left and right
rails on the cart
are included.
3
Lower drawer
assembly service
part
115-066842-00 /
The drawer
assembly and
the left and right
rails on the cart
are included.
4
Rail of the drawer
031-000664-00
/
/
4

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
BrandMindray
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals