8-32
8.35 Disassemble the Module Rack Assembly
8.35.1 Prepare for Disassembly
8.35.1.1 Tools
During parts disassembly and replacement, the following tools may be required:
Phillips screwdriver
8.35.1.2 Preparations
Before disassembly,
Make sure that the anesthesia machine is turned off and disconnected from the A/C power
source.
Maneuver the anesthesia machine to an appropriate location and then apply the brake.
8.35.2 Remove the Module Rack Assembly
1. Remove the four screws from the module rack assembly with the Phillips screwdriver.
2. Gently pull out the module rack assembly, disconnect the cables (marked by □ in the figure),
and remove the assembly.
rear of the cart
drawer mounting position