EasyManuals Logo
Home>Mindray>Medical Equipment>A9

Mindray A9 Service Manual

Mindray A9
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #278 background imageLoading...
Page #278 background image
7-54
Failure
Description
Cause Confirmation
Confirmation
Result
Solution
the output pressure of the
secondary pressure regulator
with a manometer.
kPa (22±0.7 psi).
The output pressure
is beyond 150±5
kPa (22±0.7 psi).
Adjust or replace the
secondary pressure
regulator (refer to 11.1.4).
No AA output of
the electronic
vaporizer
1. Check whether the gas
supply is sufficient.
The gas supply
pressure is less than
280 kPa (40 psi).
Connect or replace the gas
supply.
2. Check whether any
vaporizer alarm is
generated.
A vaporizer alarm
is generated.
Refer to 7.3.6 to clear the
alarm.
3. Install the electronic
vaporizer in another slot and
try to activate it.
Anesthetic agent is
output normally.
Replace the vaporizer base
valve assembly (refer to
11.4.1).
Anesthetic agent is
not output
normally.
(1) Replace the electronic
vaporizer.
(2) If the problem persists,
replace the vaporizer base
valve assembly (refer to
11.4.1).
Inaccurate
concentration
regulating of the
electronic
vaporizer
Check whether the gas
supply is sufficient.
The gas supply
pressure is less than
280 kPa (40 psi).
Connect or replace the gas
supply.
Select Setup > Service >
Data Monitors >
Vaporizer to access the data
monitoring screen and check
the drive gas pressure of
Vaporizer 1 (or Vaporizer
2).
The drive gas
pressure is within
150±5 kPa (22±0.7
psi).
Replace the electronic
vaporizer.
The drive gas
pressure is beyond
150±5 kPa (22±0.7
psi).
Adjust or replace the
secondary pressure
regulator (refer to 11.1.4).
7.4.10 AGSS Failures
Failure
Description
Cause Confirmation
Confirmation
Result
Solution
No response from
the float when the
knob is adjusted
during ventilation
with AGSS ON
Check whether the AGSS
accessories are connected
properly.
AGSS accessories
are not connected
properly.
Connect the AGSS
accessories properly.
Check whether the
hospital's waste gas
scavenging system is
normal.
The hospital's
waste gas
scavenging system
is abnormal.
Properly connect to the
waste gas scavenging
system.
Remove the AGSS
assembly by referring to
8.34 and check whether
the sampling line of the
AGSS sensor falls off.
The sampling line
of the AGSS
sensor falls off.
Reconnect the sampling
line of the AGSS sensor.
The sampling line
of the AGSS
sensor does not
fall off.
(1) Reconnect the AGSS
cable.
(2) If the problem persists,
replace the AGSS
assembly and cable (refer
to 11.1.14).

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
BrandMindray
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals