- 112 -
• code page - 1250
• prefix - 1B742D
or
• port - COM 1 or COM 2 (the one to which the printer is connected)
• code page - 852
• prefix - 1B7412
If in the place of the last digit there are any unexpected signs on the printout (for verified balances),
than <CONTROL CODES> parameter should incorporate, apart from code page, code of the UK
signs chart: 1B5203. In such a case the <CONTROL CODES> parameter setting should be as
follows: control codes - 1B74121B5203.
Control codes for exemplary code pages and functions of the printer:
25.3. RECORD OF MEASUREMENT DATA ONTO A USB FLASH DRIVE
The balance software enables record of measurement data on an external flash drive.
To perform the operation:
• Plug a USB flash drive into USB 1 port, type A.
• Set <PENDRIVE> option for <PERIPHERALS/PRINTER/PORT>.
• Return to weighing procedure.
• key, when pressed, provides record of measurement data (specified for GLP PRINTOUT),
wherein the data is saved in a form of a text file, created automatically by balance software; the
file name is: printout.txt.
• Remove the USB flash drive from a balance USB port after about 10s counting from the last
measurement record, only after this time the data is saved. Next plug the USB flash drive to
a computer and read the saved text file.