- 52 -
10.2. EXTERNAL ADJUSTMENT
External adjustment is carried out by means of an external mass standard of specified accuracy class
and weight. Both, accuracy class and mass standard weight depend on balance type and max
capacity. The process takes semi-automatic form, successive stages are signalled with prompts.
CAUTION!
External adjustment is possible for balances that are not a subject to conformity assessment (verification).
Process stages:
Enter <Adjustment> submenu, next select “External adjustment” option,
Remove weight” prompt is displayed.
Take the weight off the weighing pan and press button. Whereas balance determines start
mass, “Adjustment; Please wait…” prompt is displayed,
Upon completed start mass determination procedure “Put weight …” prompt is displayed along
with particular mass standard value.
Put the required weigh on a pan and press button,
Upon completed procedure Remove weight” prompt is displayed.
Take the weight off the weighing pan, wait for <Adjustment> window to be displayed again.
10.3. USER ADJUSTMENT
User adjustment is carried out with an optional standard of mass ranging between 0,3 Max and Max.
User adjustment and external adjustment procedures are likewise with one exception, before user
adjustment start, a message box for entering mass of a standard used for user adjustment is opened.
CAUTION!
User adjustment is possible for balances that are not a subject to conformity assessment (verification).
In order to start user adjustment, enter <Adjustment> submenu and select ‘User adjustment’ option.
Then follow the commands displayed on a screen.
10.4. ADJUSTMENT TEST
<Adjustment test> function enables comparing the result of an internal automatic adjustment with the
value recorded in balance factory parameters. The comparison is used for determining balance
sensitivity drift over time.
10.5. AUTOMATIC ADJUSTMENT
This menu is for declaring a value initiating an automatic adjustment. Accessible options:
None automatic adjustment disabled,
Time adjustment takes place in time intervals declared in menu <Automatic adjustment time>
(10.6),
Temperature adjustment is triggered by temperature change only,
Both adjustment is triggered by both, temperature changes and time.
CAUTION!
Parameter settings can be modified only for balances that are not a subject to conformity assessment
(verification).