- 64 -
ACAI, Automatic Accuracy Correction, provides the following options:
YES, single part mass is updated,
NO, single part mass is not updated.
Means of operation of ACAI function:
1. Number of pieces (on adding) on balance weighing pan has to be greater than before.
2. Number of pieces (on adding) on balance weighing pan must be less than twice the amount
displayed before adding the pieces.
3. Current quantity of parts must be within ± 0,3 tolerance of the total value.
4. Measurement result has to be stabilised.
For means of other functions operation read section 8.10 „WEIGHING mode settings”.
13.2. PARTS COUNTINGQUICK ACCESS KEYS
Each working mode features automatically displayed set of default keys. The set can be modified by
assigning out-of-list quick access keys to on-screen keys. Such operation requires particular
permissions level.
Read section 7.6.1 for list of quick access keys.
13.3. SETTING REFERENCE MASS BY ENTERING DETERMINED PART MASS
Procedure:
Press < Set part mass> key or the label , <Part mass> editing window
featuring an on-screen keyboard is displayed,