Page #66 background imageLoading...
Page #66 background image
- 66 -
The software automatically recalculates single part mass, next it moves to <Parts counting>
mode displaying number of parts loaded on the weighing pan (pcs marking).
CAUTION!
It is important to remember the following:
Total mass of all parts loaded on the weighing pan must not exceed the maximum capacity (weighing
range) of the balance;
Total mass of all parts loaded on the weighing pan must not be lower than value declared in parameter
“Minimal reference mass. Unless this condition is met, the balance displays a message: <Sample
mass too low>;
Mass of a single part must not be lower than 0,1 of balance reading unit. Unless this condition is met,
the balance displays a message: <Single part mass to low>.
13.5. ACQUIRING PART MASS FROM A DATABASE
Each product recorded in a database of products features set of information which allow to identify it.
Among the information there is mass value used during part counting process.
Procedure:
Operating <Parts counting> mode, press <
Product> button and using the list select
a particular product. The program returns to <Parts counting> mode with single part weigh value set
automatically on the basis of data recorded for a selected product.