immersed in the liquid (measurement carried out in the liquid), confirm the indication upon
stabilization.
10. Upon confirmation of the second indication, the software automatically calculates measured
liquid density, which is next displayed together with a respective message, report on
measurement is sent to a selected printer port.
An exemplary report:
The report can be printed repeatedly, press <
> key.
To finish density determination process, press key. The home screen for liquid density mode is
displayed. The next measurement can be started now. The balance operates with previously entered
settings (sinker volume) thus providing quicker start of the very measurement.
19. ANIMAL WEIGHING
< Animal weighing> is a working mode allowing reliable determination of mass of weighed
objects in motion. In principle, this type of objects generate unstable measurement, thus it requires
using a different filtering method of measurement signal.
Procedure for activation of Animal Weighing mode