Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-49
:MASK:OUTPut
Syntax :MASK:OUTPut <bool>
Description Enable or disable the sound prompt when the failed waveforms are detected, or query the
status of the sound prompt.
Parameter
Name Type Range Default
<bool>
{{0|OFF}|{1|ON}}
Explanation  OFF: when failed waveforms are detected, there are display and output but the
beeper does not sound.
 ON: when failed waveforms are detected, there are display and output and the
beeper sounds (not related to the on/off state of the sound).
Return
The query returns 0 or 1.
:MASK:X
Syntax :MASK:X <x>
Set or query the horizontal adjustment parameter in the pass/fail test mask.
Parameter
<x> Real 0.02 to 4 0.24
The step is 0.02 within the range.
Return
The query returns the horizontal adjustment parameter in scientific notation.
Example
:MASK:X 0.28 /*Set the horizontal adjustment parameter to 0.28div*/
:MASK:X? /*The query returns 2.800000e-01*/
:MASK:Y
Syntax :MASK:Y <y>
Set or query the vertical adjustment parameter in the pass/fail test mask.
Parameter
The step is 0.04 within the range.
Return
The query returns the vertical adjustment parameter in scientific notation.
Example
:MASK:Y 0.36 /*Set the vertical adjustment parameter to 0.36div*/
:MASK:Y? /*The query returns 3.600000e-01*/