RIGOL Chapter 2 Command System
2-56 DS1000Z Programming Guide
:MEASure:ADISplay
Syntax :MEASure:ADISplay <bool>
Description Enable or disable the all measurement function, or query the status of the all
measurement function.
Parameter
Explanation The all measurement function can measure the following 20 parameters of the
source at the same time:
Voltage Parameters: Vmax, Vmin, Vpp, Vtop, Vbase, Vamp, Vavg, Vrms, Overshoot
and Peshoot
Time Parameters: Period, Frequency, Rise Time, Fall Time, + Width, - Width, + Duty
and - Duty
Other Parameters: Area and Period Area.
The all measurement function can measure CH1, CH2, CH3 and CH4 at the same
time. You can send the
:MEASure:AMSource command to set the source of the all
Return
The query returns 0 or 1.
Related
:MEASure:AMSource
:MEASure:AMSource
Syntax :MEASure:AMSource <src>[,<src>[,<src>[,<src>]]]
Set or query the source of the all measurement function.
Parameter
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
Return
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.
Example