EasyManuals Logo
Home>Rigol>Test Equipment>DS1000Z Series

Rigol DS1000Z Series Programming Guide

Rigol DS1000Z Series
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #70 background imageLoading...
Page #70 background image
RIGOL Chapter 2 Command System
2-56 DS1000Z Programming Guide
:MEASure:ADISplay
Syntax :MEASure:ADISplay <bool>
:MEASure:ADISplay?
Description Enable or disable the all measurement function, or query the status of the all
measurement function.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{0|OFF}|{1|ON}}
0|OFF
Explanation The all measurement function can measure the following 20 parameters of the
source at the same time:
Voltage Parameters: Vmax, Vmin, Vpp, Vtop, Vbase, Vamp, Vavg, Vrms, Overshoot
and Peshoot
Time Parameters: Period, Frequency, Rise Time, Fall Time, + Width, - Width, + Duty
and - Duty
Other Parameters: Area and Period Area.
The all measurement function can measure CH1, CH2, CH3 and CH4 at the same
time. You can send the
:MEASure:AMSource command to set the source of the all
measurement function.
Return
Format
The query returns 0 or 1.
Example
:MEASure:ADISplay ON
Related
Command
:MEASure:AMSource
:MEASure:AMSource
Syntax :MEASure:AMSource <src>[,<src>[,<src>[,<src>]]]
:MEASure:AMSource?
Description
Set or query the source of the all measurement function.
Parameter
Name
Type
Range
Default
<src>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
Return
Format
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.
Example
:MEASure:AMSource CHANnel1,CHANnel3

Table of Contents

Other manuals for Rigol DS1000Z Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000Z Series and is the answer not in the manual?

Rigol DS1000Z Series Specifications

General IconGeneral
BrandRigol
ModelDS1000Z Series
CategoryTest Equipment
LanguageEnglish

Related product manuals