Performance Verification
4-2
TLA5000 Series Service Manual
Determine equipment
to certify.
Go to appropriate
procedure(s).
Gather required test
equipment.
Use
Tektronix-supplied
probes.
Configure customer’s logic
analyzer as described in
procedure.
Set up test equipment.
Select procedure.
Allow the logic analyzer and test
equipment to warm up (30 minutes ).
Are
all certifications
complete?
Done
Is it OK
with customer to
remove/reconnect probes
from System
Under Test?
No
No
Yes
Do tests
pass?
Repair or adjust
equipment as
required.
No
Yes
Yes
Perform procedure
using software.
Load PV/Adjust
software.
Create necessary
calibration certificate
from data stored in
certification file.
Figure 4- 1: Calibration/certification procedure flow chart