Performance Verification
4-14
TLA5000 Series Service Manual
6. Verify that Pulse Threshold is selected under Module+Probe Gain & Offset
Procedures. Clear all other procedure selections in the dialog box.
NOTE. The software will perform all selected procedures without interruption,
with the exception of when user interaction is required (such as moving a probe).
To limit the test to specific procedures, clear any procedures that you do not
want to run. If you want to run all of the selected procedures, you need to pay
close attention to the connection instructions as they are displayed on the screen.
7. Click the Next button at the bottom of the dialog box to display the probe
connection instructions. Follow the on-screen instructions to connect the
probes from the logic analyzer to the test fixture. If necessary, refer to the
test fixture label to determine the correct probe connection.
NOTE. When connecting the P6419 Logic Analyzer probes, make sure that the
alignment pin on the probe head aligns w ith the hole on the test fixture. Tighten
the probe head screws by alternating between them until they are finger tight (no
more than 1 in-lbs of torque).
8. Click the Next button to begin the verification procedure.
The software will begin the procedures and display a list of results in the
window.
9. After the first set of procedures are done, click the Next button to display the
instructions for the next step. In most cases you only need to connect a
different probe to the test fixture.
10. After changing the connections, click the Next button to continue the tests.
11. Repeat steps 7 through 8 as indicated by the software.
12. After the procedure finishes, click the Next button to open the Finish dialog
box where you can save the results to a file. Enter a file name and click the
Save button. You can use the Save dialog to save the file to a folder of your
choice or use the default location.
13. Click the Finish button to complete the process.