XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30 52
All the following information is the property of Dürr dental. No data or indications have been modified.
Those informative have been added to this guide to provide the necessary information to be able to
quickly intervene on a failure related to the Vistacam iX or iX HD camera.
For more information, please refer to the "Vistacam iX or iX HD - Installation and Operating Instructions"
at http://www.durrdental.com/
The hygienic protective cover is
not placed correctly on the
Place the hygienic protective
cover on the optical window
Optical element scratched
Replace the interchangeable
head.
The handpiece is defective
Send the handpiece for repair
Lens above image sensor dirty
Replace the interchangeable
head
USB connection cable not
connected
Connect the USB connection
cable
USB connection cable too long
For long connection to the clinic
PC, use an USB repeater
Computer not switched on;
software not started
Switch on the computer and start
the software.
Camera driver not correctly
installed
Check the driver installation and
software settings.
Interchangeable head not placed
on correctly, no contact between
the handpiece and the
interchangeable head
Ensure that the interchangeable
head has been placed on to its
fullest extent, no gap between
the handpiece and the
pressing the trigger button,
but no picture is displayed
recognized by software
Replace the interchangeable
head.
computer is too low
Use a computer in accordance
with the system requirements
Resolution not set correctly
IX: 1024x768