XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30 83
SPITTOON VALVE DÜRR
All the following information is the property of Dürr dental. No data or indications have been
modified and have only been added to this guide to provide the necessary information to be
able to quickly intervene on a failure related to the spittoon valve.
For more information, please refer to the "Installation and Operating Instructions" at
http://www.durrdental.com/
Failure description
Reason
Solution
Spittoon valve is not
working.
No Power supply
Verify the power supply
Check the connectors
No air present
Check the compressed air
supply on spittoon valve.
Sensor is defective
Check the sensor function by
pressing the switch.
Check the function by manually
moving the float sensor.
Suction unit does not start
or run continuously.
The float sensor is blocked
Clean the housing and the float
sensor.
Insert the float sensor correctly.
The liquid does not flow
through
The drain is blocked
Clean the drain line.
Check if the filters. Clean them if
clogged