4263B
T
o Apply
aDC
Bias
The
following commands
are
used
to
apply
a
dc
bias
voltage:
:SOUR:VOLT:OFFS
:SOUR:VOLT:OFFS:SOUR
:SOUR:VOLT:OFFS:STAT
F
or
example
,
to
apply
1.5
V
DC
bias
using
the
internal
bias
source,
OUTPUT
717;":SOUR:VOLT:OFFS:SOUR
INT"
OUTPUT
717;":SOUR:VOLT:OFFS
1.5"
OUTPUT
717;":SOUR:VOLT:OFFS:STAT
ON"
:
OUTPUT
717;":SOUR:VOLT:OFFS:STAT
OFF"
T
o
P
erform
Correction
The
following
commands
perform
the
correction
function:
:SENS:CORR
:SENS:CORR:COLL
:SENS:CORR:COLL:METH
:SENS:CORR:CKIT:STAN3
There
are
two
correction
methods
as
follows:
OPEN
and
SHORT
correction
T
o
correct
for
stray
admittance
and
residual
impedance
due
to
the
test
xture.
OPEN,
SHORT
,
and
LO
AD
correction
T
o
correct
for
any
error
due to
the
test
xture
and
test
leads
by
using
a
standard
(LO
AD).
Refer
to
\Correction
Functions
of
the
4263B
"in
Chapter
7
for
more
information.
Note
When
the
measurement
parameter
is
Ls-Rdc
,
Lp-Rdc
, L2-N,
or L2-R2,
the
LO
AD
correction can
correct only
the
errors
for
Ls
,
Lp
,
or
L2.
4-8 Remote Operation (To Control from a Computer)

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals