4263B
Handler
Interface
Functional
T
est
The
4263B
's
handler
interface
function
is
tested
using
the
built-in self-test
and the
handler
interface
tester
.
T
est
Equipment
Description
Recommended Model
Handler
Interface T
ester p/n
04339-65007
Procedure
Initial
Setup
.
1.
Turn
the
4263B
OFF.
2. Set
the LED
PW switch
on the
Handler Interface
T
ester
to
OFF
,
and
set
the
IN1,
IN2,
IN3,
and IN4
switches to
0.
3.
Set
up
the
equipment
as
shown
in Figure
9-10
Figure
9-10.
Handler
Interface
Functional
T
est
Setup
4.
Turn
the
4263B
ON.
5.
Reset the
4263B.
K
ey Lock
Function T
est.
6.
Set
the
IN4
switch
on
the
Handler
Interface
Tester
to
1.
7. Conrm that the all keys on the 4263B's front panel are locked out.
8. Set the
IN4 switch on the Handler Interface T
ester to 0.
External Trigger Function T
est.
9. Press
to set the trigger mode to
External.
10. Press the IN5 switch on the Handler Interface Tester, and conrm that the 4263B is
triggered.
Maintenance 9-49

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals