C
Overload/No-Contact
Operations
Table
C-1
shows
the
summary
of
operations
,
when
the
4263B
detects
O
VLD
(Overload),
or
N.C.
(No-Contact).
Table
C-1.
O
VLD/N.C.
Condition
Display Handler
Output
GPIB
Output
1
Solutions
Data Mode Comprtr
Mode
O
VLD
(Overload)
O
VLD
High
4
/PHI
and
/SHI
<
State
>
2
:1
<
Data
>
:9.9E37
<
Comp
>
3
:2
Select
an
approptiate
measurement
range
.
Low
4
/PLO
and
/SLO
<
State
>
2
:1
<
Data
>
:9.9E37
<
Comp
>
3
:4
N.C.
(No-Contact)
N.C. N.C. /NO
CONT
A
CT
<
State
>
2
:2
<
Data
>
:9.9E37
<
Comp
>
3
:8
Cancel
the
bad
contact
between
the
DUT
and
the
contact-pin.
1.
Refer
to
the
:FETCh?
Command
in
Chapter
5.
2.
Measurement
Status
:
(0
:
Normal,
1
:
Overload,
2
:
No-Contact)
3.
Comparison
result
:
(1
:
IN,
2
:
HIGH,
4
:
LOW
,
8
:
No-Contact)
4.
Comparison
display
:
Depends
on
the
relation
between
the 4263B
measurement range
and
the
impedance
of
the
DUT
.
Overload/No-Contact Operations C-1

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals