4263B
1
mDC
Resistance Measurement
A
ccuracy
T
est
(Opt.
001
Only).
22.
Set the
4263B
measurement
condition
as
follows
.
Measurement
Parameter:
Ls-Rdc
Measurement Time
Mode:
LONG
T
est
Signal
Level:
1
V
Trigger
Mode:
Man
23.
Connect
the
100
m
Standard
Resistor
to
the
4263B
's
UNKNOWN
terminals
using
the
1
m
test
leads
and
four
BNC(f)-BNC(f)
adapters
.
24.
Press
to
start
the
measurement.
25.
Record
the
4263B
readings
of
Rdc
on
the
calculation
sheet.
26.
Calculate
the
test
results
according
to
the
calculation
sheet,
and
record
the
result
into
the
performance
test
record.
27.
Perform
this test
for the
100 k
standard
resistor
.
Maintenance 9-21

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals