4263B
29. Set
the DC
bias voltage
to
2
V
according
to
the
following
procedure:
a. Press
to display
the DC
Bias Setting.
b.
Press
until
2
V
blinks
,
and
press
.
c.
Press
to
turn
the
DC
Bias
ON.
(DC
Bias indicator
ON.)
30. Press
to start
the measurement.
31.
Record
the
4263B
readings
of
Cp
and
D
on
the
calculation
sheet.
32.
Calculate
the
test
results
according
to
the
calculation
sheet,
and
record
the
results
into
the
performance
test
record.
33.
Press
to
turn
the
DC
Bias
o.
T
able
9-8.
0
m
Capacitance
Measurement
Test
(DC Bias)
Setting
Standard Test
Signal
Frequency
1

F
1
kHz
0
m
Resistance
Measurement
A
ccuracy
T
est
(Meas
.
Time
Mode:
LONG).
34.
Set
the
4263B
measurement
conditions
as
follows:
Measurement
P
arameter:
R-X
Measurement
Time
Mode:
LONG
T
est Signal
Level:
1
V
T
est
Signal
Frequency:
100
Hz
Trigger
Mode:
Man
35. Connect
the 100
m Standard
Resistor to
the 4263B
's UNKNOWN
terminals.
36. Press
to start the measurement.
37. Record the 4263B readings of R on the calculation sheet.
38. Calculate the test results according to the
calculation sheet, and record the result into the
performance test record.
39. P
erform the test at 1 kHz.
Maintenance 9-15

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals