4263B
Front-end Protection
Internal input
circuit
is
protected,
when
a
charged
capacitor
is
connected
to
the
Input
terminals.
The maximum
capacitor
voltage
is:
1000 V
(C
<
2

F)
p
2
=C
V
(2

F

C
<
32

F)
250
V
(32

F

C
<
128

F)
p
8
=C
V
(C

128

F)
(where
,
C
is
the
capacitance
value
of
capacitor
(F)
)
Figure
8-2.
Maximum
Capacitor
Voltage
Caution
Do
not
apply
DC
voltage
or
current
to
the
UNKNOWN
terminals
.
Doing
so will
damage
the
4263B
.
Before
you
measure
a
capacitor
,
be
sure
the
capacitor is
fully
discharged.
Recall
time of
the saved
instrument
setup
approx.
200 ms