EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>4263B

Agilent Technologies 4263B User Manual

Agilent Technologies 4263B
330 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #219 background imageLoading...
Page #219 background image
4263B
Selecting
Circuit
Model
for
Capacitance
Measurement
The
following
description
gives
some
practical
guide
lines
for
selecting
the appropriate
capacitance
measurement
circuit
model.
Small
Capacitance
V
alues
Small
capacitance
yields
a
large
reactance
,
which
implies
that
the eect
of the
large value
parallel
resistance
(Rp)
has
relatively
more
signicance than
that of
the low
value series
resistance
(Rs).
So
the
parallel
circuit
model (Cp-D
, Cp-Q,
or Cp-G)
is the
one
to
use
.
(Refer
to
Figure
7-2
(a).)
Large
Capacitance
V
alues
When
the
measurement
involves
a large
value of
capacitance (low
reactance),
Rs
has
relatively
more
signicance
than
Rp
,so
the series
circuit model
(Cs-D
,
Cs-Q,
or
Cs-Rs)
is
the
one
to
use
.
(Refer
to
Figure
7-2
(b).)
Figure
7-2.
Capacitance
Circuit
Model
Selection
The
following
are
rules
of
thumb
for
selecting
the
circuit
model according
to the
reactance
of
the
capacitor
.
Above
approx.
10
k
:
Use
parallel
circuit model
Below
approx. 10
:
Use series
circuit
model
Between
above values
: F
ollow the
manufacturer's
recommendation
F
or
example,
to measure
a20
nF capacitor
at
100
Hz
(reactance
is
approximately
80
k),
the
parallel circuit
model is
suitable.
7-4 Impedance Measurement Basics

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals