Agilent 4263B LCR Meter Operation Manual
Manual Change
Agilent Part No. N/A
Jun 2009
Change 1
Add TAR in Test Signal Frequency Accuracy Test (Page 9-38) as follows.
Test Signal Frequency Accuracy
Frequency
[Hz]
Limits [Hz] Result
1
[Hz] Uncertainty [Hz] TAR
100 ±0.01 ± 0.00031 31.77
120 ±1.2 ± 0.00034 3550.50
1 k ±0.00010 k k ± 0.0000029 k 35.38
10 k ±0.0010 k k ± 0.000019 k 53.80
20 k ±0.0020 k k ± 0.000020 k 100.32
100 k ±0.010 k k ± 0.00019 k 53.78
Note 1: Result = Measurement – Nominal Frequency
Change 2
Add TAR in Test Signal Level Accuracy Test (Page 9-38) as follows.
Test Signal Level Accuracy
Signal
Level [V]
Frequency
[HZ]
Limits [V] Result
1
[V] Uncertainty [V] TAR
50 m 100 ±15.0 m
m
± 0.90 m 16.70
50 m 20 k ±15.0 m
m
± 0.078 m 193.62
50 m 100 k ±15.0 m
m
± 0.18 m 86.62
250 m 100 ±35.0 m
m
± 2.50 m 14.04
250 m 20 k ±35.0 m
m
± 0.35 m 102.30
250 m 100 k ±35.0 m
m
± 1.88 m 18.66
1 100 ±0.11 ± 0.00349 31.60
1 20 k ±0.11 ± 0.00147 74.91
1 100 k ±0.11 ± 0.00108 101.92
Note 1: Result = Measurement – Nominal Level
Change 3
Add TAR in DC Bias Level Accuracy Test (Page 9-39) as follows.
DC Bias Level Accuracy
DC Bias
Level [V]
Limits [V] Result1 [V] Uncertainty [V] TAR
0 ±0.002 ± 0.087 m 23.07
1.5 ±0.077 ± 0.44 m 175.51
2 ±0.102 ± 0.92 m 111.10
Note 1: Result = Measurement – Nominal Level

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals