â—‹
C
Copyright 2009 Agilent Technologies
Change 4
Add TAR in 0 m Impedance Measurement Accuracy Test (Page 9-40 to 9-42) as follows.
Impedance Measurement Accuracy
Measurement:
Capacitance
Cable length:
0 m
Measurement time:
Long
Test signal level:
1 V
Status
PASS
Standard
Frequency [Hz]
Parameter Limits Result
1
Uncertainty
TAR
10 pF 100 k Cp ±0.331 pF
pF
± 0.0040 pF
83.58
10 pF 100 k D ±0.0331 ± 0.00043
78.09
100 pF 1 k Cp ±0.19 pF
pF
± 0.0074 pF
25.79
100 pF 1 k D ±0.0019 ± 0.000063
30.57
100 pF 10 k Cp ±0.44 pF
pF
± 0.017 pF
26.79
100 pF 10 k D ±0.0044 ± 0.00023
19.15
100 pF 20 k Cp ±1.10 pF
pF
± 0.021 pF
53.34
100 pF 20 k D ±0.0110 ± 0.000091
121.51
100 pF 100 k Cp ±1.37 pF
pF
± 0.019 pF
74.59
100 pF 100 k D ±0.0137 ± 0.00033
42.12
1 nF 100 Cp ±0.002 nF
nF
± 0.000098 nF
20.11
1 nF 100 D ±0.0020 ± 0.000063
32.10
1 nF 1 k Cp ±0.0012 nF
nF
± 0.000092 nF
12.67
1 nF 1 k D ±0.0012 ± 0.000064
19.02
1 nF 100 k Cp ±0.0128 nF
nF
± 0.00012 nF
116.20
1 nF 100 k D ±0.0128 ± 0.00033
38.90
10 nF 100 Cp ± 0.018 nF
nF
± 0.00065 nF
28.46
10 nF 100 D ±0.0018 ± 0.000022
84.00
10 nF 120 Cp ± 0.018 nF
nF
± 0.00059 nF
16.38
10 nF 120 D ±0.0018 ± 0.000022
85.34
10 nF 1 k Cp ± 0.011 nF
nF
± 0.00076 nF
14.17
10 nF 1 k D ±0.0011 ± 0.000021
53.05
10 nF 10 k Cp ± 0.018 nF
nF
± 0.00067 nF
27.75
10 nF 10 k D ±0.0018 ± 0.000036
51.18
10 nF 100 k Cp ± 0.128 nF
nF
± 0.00087 nF
148.58
10 nF 100 k D ±0.0128 ± 0.00022
60.42
100 nF 1 k Cp ± 0.11 nF
nF
± 0.0068 nF
15.62
100 nF 1 k D ±0.0011 ± 0.000023
47.81
100 nF 100 k Cp ± 1.47 nF
nF
± 0.015 nF
102.37
100 nF 100 k D ±0.0147 ± 0.00026
57.89
1 μF 100 Cp ±0.0018 μF
μF
± 0.000074 μF
24.88
1 μF 100 D ±0.0018 ± 0.000047
38.90
1 μF 120 Cp ±0.0018 μF
μF
± 0.000073 μF
24.49
1 μF 120 D ±0.0018 ± 0.000047
38.84
1 μF 1 k Cp ±0.0011 μF
μF
± 0.000067 μF
15.89
1 μF 1 k D ±0.0011 ± 0.000032
34.29
1 μF 10 k Cp ±0.0026 μF
μF
± 0.00014 μF
19.45
1 μF 10 k D ±0.0026 ± 0.000069
38.02
1 μF 100 k Cp ±0.0176 μF
μF
± 0.00018 μF
102.72