EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>4263B

Agilent Technologies 4263B User Manual

Agilent Technologies 4263B
330 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #258 background imageLoading...
Page #258 background image
4263B
0
m
Impedance
Measurement
A
ccuracy
T
est
The
4263B
measures
the
calibrated
standard
capacitors
and
resistors
at
the 4263B
's front
panel,
and
the
measured
values
are
compared
with the
standards' listed
calibration values
.
Specication
Basic
Measurement
A
ccuracy:
6
0.1
%
(See
Chapter
8
General Information
for
details
.)
T
est
Equipment
Description
Recommended Model
Standard
Capacitor
Set
16380A
Standard
Capacitor
Set
16380C
100
m
Standard
Resistor
42033A
1
100
k
Standard
Resistor
42039A
1
,
2
OPEN
T
ermination 42090A
2
SHORT
T
ermination
42091A
2
1
P
art
of
42030A
Standard
Resistor
Set
2
16074A
can
be
used
as
a
substitute
.
Procedure
0
m
Capacitance
Measurement
A
ccuracy
T
est
(Meas
.
Time
Mode:
LONG).
1.
Record
the
16380A,
16380C,
and
42030A
calibration
values
on
the calculation
sheet.
2.
Reset
the
4263B
.
3.
Connect
the
OPEN
termination
to
the
4263B
's
UNKNOWN
terminals
.
Figure 9-5. 0 m Impedance Measurement Accuracy Test Setup
Maintenance 9-11

Table of Contents

Other manuals for Agilent Technologies 4263B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4263B and is the answer not in the manual?

Agilent Technologies 4263B Specifications

General IconGeneral
BrandAgilent Technologies
Model4263B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals