User manual SEFELEC 5x Series 181 v1.06
18.1.4. COMMANDS DEPENDENT ON THE DEVICE
MEG (MEGohmmeter) :
Selects the megohmmeter. Must be sent from the initialization display.
HIP (HIPot) :
Selects the Hipot function. Must be sent from the initialization display.
GND (GrouND) :
Selects ground continuity function. Must be sent from the initialization display.
SEQ (SEQuence) :
Selects the sequence function. Must be sent from the initialization display.
CONF (CONFig) :
Selects the configuration function of the unit. Must be sent from the initialization
display.
PAR (PARameter) <NR1> :
Selects the parameter set for the running function.
i.e.: MEG : PAR 1
DCV (DCVoltage) <NR1> :
Sets a new value in volt to the current DC voltage parameter to the parameter set
of the running function.
i.e. : DCV 500
ACV (ACVoltage) <NR1> :
Sets a new value in volt to the current AC voltage parameter to the parameter set
of the running function without changing the pre-set frequency.
Hipot tester : Min. : 100 Max. : 5000
For instance : hipot : ACV 5000
ACV50 (ACVoltage) <NR1> :
Sets a new value in volt to the current AC (50Hz frequency) voltage parameter to
the parameter set of the running function
Hipot tester : Min. : 100 Max. : 5000
For instance : hipot : ACV50 5000
ACV60 (ACVoltage) <NR1> :
Sets a new value in volt to the current AC (60Hz frequency) voltage parameter to
the parameter set of the running function
Hipot tester: Min.: 100 Max. : 5000
For instance: hipot : ACV60 5000
ACC (ACCurrent) <NR3> :
Sets a new value in ampere to the AC current parameter to the parameter set of
the ground continuity function.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eaton SEFELEC 56-D and is the answer not in the manual?

Eaton SEFELEC 56-D Specifications

General IconGeneral
BrandEaton
ModelSEFELEC 56-D
CategoryTest Equipment
LanguageEnglish

Related product manuals