User manual SEFELEC 5x Series 66 v1.06
SELECTION OF THE FILTER MODE
The device offers a filter mode to stabilize the measurement on the capacitive components
Press the active area of the "Filter" line activate the filtering mode:
The device offers a measurement configuration for capacitive circuits:
Capacitor: this measurement mode implements analogue treatments and software to provide
stable measurements on highly capacitive circuits.
The activation of the Capacitor mode is materialised by the symbol of a capacitor in the top left
of the main measurement window.
Capacity mode icon:
It is imperative to switch to discharge with the Stop button before disconnecting the component
under test.