User manual SEFELEC 5x Series 183 v1.06
FILT (FILTer) NOR/CAP
Enables the normal or the capacitor measurement modes.
For instance : CONF:FILT CAP
DET (DETection) OFF / I / I+DELTA / DELTA :
Hipot function only sets the mode of the breakdown.
IDE (IDElta) <NR3>
Hipot function : Delta I breakdown detection value adjustment
MKM (MEGOHMKM) ON OFF
Allows to activate or disactivate the Megohm.km mode
MKMLEN (MEGOHMKMLENGTH) <NR1>
with <NR1> cable length (1 to 16000 meters) in Megohm.km mode
MKMCOUNT (MEGOHMKMCOUNT) <NR1>
With <NR1> number of wires (0 to 999) in Megohm.km mode
MEAS (MEASure)
Runs the current function.
STOP
Stops the current function.
QUIT
Exits the current function.
SRQ
Similar to the Service Request feature of the IEEE488-2 bus, this code allows the
emission of the Z character to inform the computer about the events (end of test,
format error, interlock open, ). This command has to be sent at the program
start, after the REM command.
18.1.5. REQUEST DEPENDING ON THE DEVICE
MEAS?
Return the current measurement value(s) regarding the running function.
Insulation : VOLT 9.900E+01 AMP 2.000E-06 OHM 4.950E+06
Dielectric strength : VOLT 1.000E+03 AMP 7.000E-05 AMP 0.100E-06
Continuity : VOLT 2.810E+00 AMP 1.000E+01 OHM 2.810E-01