113
Product Certifications
Product Certifications
November 2016
Reference Manual
00809-0100-4804, Rev CB
Appendix B Product Certifications
Rev. 1.9
European Directive Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 113
Ordinary Location Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 113
Installing Equipment in North America . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 113
USA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 113
Canada . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 114
Europe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 114
International . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 115
Brazil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 116
China . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 117
EAC — Belarus, Kazakhstan, Russia . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 119
Japan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 119
Republic of Korea . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 119
Combinations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 119
Additional Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 119
B.1 European Directive Information
A copy of the EU Declaration of Conformity can be found at the end
of the Quick Start Guide. The most recent revision of the EU
Declaration of Conformity can be found at
Emerson.com/Rosemount
.
B.2 Ordinary Location Certification
As standard, the transmitter has been examined and tested to
determine that the design meets the basic electrical, mechanical,
and fire protection requirements by a nationally recognized test
laboratory (NRTL) as accredited by the Federal Occupational Safety
and Health Administration (OSHA).
B.3 Installing Equipment in North
America
The US National Electrical Code (NEC) and the Canadian Electrical
Code
®
(CEC) permit the use of Division marked equipment in
Zones and Zone marked equipment in Divisions. The markings
must be suitable for the area classification, gas, and temperature
class. This information is clearly defined in the respective codes.
B.4 USA
E5 FM Explosionproof (XP) and Dust-Ignitionproof (DIP)
Certificate: 3008216
Standards: FM Class 3600 – 2011, FM Class 3615 – 2006, FM
Class 3616 – 2011, FM Class 3810 – 2005,
ANSI/NEMA 250 – 2003
Markings: XP CL I, DIV 1, GP B, C, D; DIP CL II, DIV 1, GP E, F, G;
CL III; T5(–50 °C
Ta +85 °C); Factory Sealed; Type
4X
I5 FM Intrinsic Safety (IS) and Nonincendive (NI)
Certificate: 3012350
Standards: FM Class 3600 – 2011, FM Class 3610 – 2010, FM
Class 3611 – 2004, FM Class 3810 – 2005, NEMA
®
250 – 2003
Markings: IS CL I, DIV 1, GP A, B, C, D; CL II, DIV 1, GP E, F, G;
Class III; Class 1, Zone 0 AEx ia IIC T4; NI CL 1, DIV
2, GP A, B, C, D; T4(–50 °C
Ta +70 °C) [HART];
T4(–50 °C Ta +60 °C) [Fieldbus]; when
connected per Rosemount drawing 03151-1006;
Type 4x
Special Condition for Safe Use (X):
1. The Rosemount
3051S/3051S-ERS Pressure Transmitter
contains aluminum and is considered to constitute a
potential risk of ignition by impact or friction. Care must be
taken into account during installation and use to prevent
impact and friction.
Note
Transmitters marked with NI CL 1, DIV 2 can be installed in Division
2 locations using general Division 2 wiring methods or
Nonincendive Field Wiring (NIFW). See Drawing 03151-1006.
IE FM FISCO
Certificate: 3012350
Standards: FM Class 3600 – 2011, FM Class 3610 – 2010, FM
Class 3611 – 2004, FM Class 3810– 2005, NEMA
250 – 2003
Markings: IS CL I, DIV 1, GP A, B, C, D; (–50 °C
Ta +60 °C);
when connected per Rosemount drawing
03151-1006; Type 4x

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount 3051S ERS and is the answer not in the manual?

Emerson Rosemount 3051S ERS Specifications

General IconGeneral
Model3051S ERS
CategoryAccessories
Operating Temperature-40 to 85°C (-40 to 185°F)
Transmitter CompatibilityRosemount 3051S
Communication ProtocolHART
Housing MaterialStainless Steel
CompatibilityRosemount 3051S
Enclosure RatingNEMA 4X
Product TypeElectronic Remote Sensor (ERS) System