65
Specifications and Reference Data
Specifications and Reference Data
November 2016
Reference Manual
00809-0100-4804, Rev CB
Appendix A Specifications and Reference Data
Performance specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 65
Functional specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 68
Physical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 71
Dimensional drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 73
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 82
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 109
A.1 Performance specifications
The Rosemount 3051S Electronic Remote Sensor (ERS)™ System
offer three performance class options; Classic, Ultra, and Enhanced
ERS System Performance. The Classic and Ultra performance
classes are suited to lower static pressure and stable temperature
conditions. The Enhanced ERS System Performance class provides
better performance across temperature (–40 to 185 °F) with
improved performance at higher static pressure.
A.1.1 Conformance to specification (±3)
Technology leadership, advanced manufacturing techniques, and
statistical process control ensure specification conformance to at
least ±3.
A.1.2 Reference accuracy
For zero-based spans, reference conditions, silicone oil fill,
glass-filled PTFE O-rings, SST materials, coplanar flange or
1
/2–14
NPT process connections, digital trim values set to equal range
points. Stated reference accuracy equations include terminal
based linearity, hysteresis, and repeatability.
Table A-1. DP Total Accuracy for Enhanced ERS System Performance
(1)
Sensor type
3051SAM_ _G2,
3051SAL_ _G2
250 inH
2
O
(622,1 mbar)
3051SAM_ _G3,
3051SAL_ _G3
1000 inH
2
O
(2488,4 mbar)
3051SAM_ _T1,
3051SAL_ _T1
30 psi
(2,1 bar)
3051SAM_ _T2,
3051SAL_ _T2
150 psi
(10,34 bar)
3051SAM_ _G4,
3051SAL_ _G4
300 psi
(20,7 bar)
3051SAM_ _T3,
3051SAL_ _T3
800 psi
(41,4 bar)
Rosemount 3051SAM
(2)
0.2 inH
2
O
(0,5 mbar)
0.6 inH
2
O
(1,4 mbar)
0.9 inH
2
O
(2,2 mbar)
1.5 inH
2
O
(4,0 mbar)
6.2 inH
2
O
(15 mbar)
7.8 inH
2
O
(19 mbar)
Rosemount 3051SAL with
direct mount seal types
and sizes below
(3)
:
FF, FC, PF 2-in./DN50
EF 3-in./DN80
All RT, RF, RC, SS
SC 2.5-in.
2.2 inH
2
O
(5,5 mbar)
2.3 inH
2
O
(6,0 mbar)
3.0 inH
2
O
(7,5 mbar)
3.2 inH
2
O
(8,0 mbar)
6.5 inH
2
O
(16 mbar)
8.3 inH
2
O
(21 mbar)
Rosemount 3051SAL with
other seal types and sizes
Consult Instrument Toolkit
for performance.
1. Includes full ambient and temperature range from –40 to 85 °C (–40 to 185 °F) requires two transmitters with identical sensor ranges. Specification are only
applicable for spans down to 10:1.
2. For Rosemount 3051SAM assembled to a Rosemount 1199 Diaphragm Seal, use Rosemount 3051SAL specification for identical seal types and sizes.
3. For Rosemount 3051SAL with direct mount seals, specification applies to process temperatures from -45 to 205 °C and excludes diaphragm option code SC, 6-mil
diaphragm thickness.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount 3051S ERS and is the answer not in the manual?

Emerson Rosemount 3051S ERS Specifications

General IconGeneral
Model3051S ERS
CategoryAccessories
Operating Temperature-40 to 85°C (-40 to 185°F)
Transmitter CompatibilityRosemount 3051S
Communication ProtocolHART
Housing MaterialStainless Steel
CompatibilityRosemount 3051S
Enclosure RatingNEMA 4X
Product TypeElectronic Remote Sensor (ERS) System